Spherical Chamber 10 Liter 32003

Spherical ionization chamber for radiation protection measurement

  • Vented sensitive volume of 10 liters
  • Suitable for survey meter calibration and low level measurements
  • Superior energy response, reproducibility, directional dependence and long-term stability

The spherical chamber is designed for the measurement of ionizing radiation in radiation protection.

Superior features make the chamber suitable as standard chamber for calibration purposes. It fulfills the requirement for excellent reproducibility and long-term stability of the sensitive volume.

The spherical construction ensures a nearly uniform response to radiation from every direction.
The energy response is very flat. This is achieved by the thin layer of aluminum on the inner wall surface, which provides for an increased photoelectric yield to compensate for the absorption of soft X-rays. The outer chamber diameter is 275 mm

Type of product vented spherical
  ionization chamber
Application radiation protection measurement
Measuring quantities air kerma, photon equivalent dose
Design not waterproof, vented
Nominal sensitive volume 10 l
Reference point chamber center
Nominal response 300 μC/Gy
Chamber voltage 400 V nominal, ± 500 V maximal
Energy response ≤ ± 3 %
Leakage current ≤ ± 10 fA

Materials and measures

Wall of sensitive volume 2.75 mm POM (polyoxymethylene),
  0.02 mm graphite
  0.22 mm lacquer
Total wall area density 417 mg/cm2
Central electrode graphite coated polystyrene,
  diameter 100 mm
Outer dimensions diameter 276 mm

Ion collection efficiency at nominal voltage

Ion collection time 150 ms
Max. dose rate for ≥ 90.0 % saturation 26 mGy/h
Max. dose per pulse for ≥ 99.0 % saturation 0.8 μGy

Ranges of use

Chamber voltage ± (300 ... 500) V
Radiation quality 25 keV ... 50 MeV
Temperature (10 ... 40) °C
  (50 ... 104) °F
Humidity (10 ... 80) %, max 20 g/m3
Air pressure (700 ... 1060) hPa

Options

T48010 Radioactive check device 90Sr
T48001 Chamber holding device for check device
Diese Seite existiert auch in Ihrer Sprache.
This page also exists in your language.
Sprache wechseln Change Language